Zero calibration is the process of resetting the reading so that it displyas 0 when measuring a gas that does not contain CO2.
✔ The carbon dioxide concentration in the room is normally almost zero (around 0.04%), but it measures a much greater value than zero, even though nothing is connected to it.
✔ When using the zero-point adjustment button without a zero calibrator, a patient with a value of 38 may be measured as low as 35.
✔ Also, an obvious outliner may be measured, such as over 60.
In these cases, you should perform zero calibration by using a zero calibrator.
That completes the zero calibration process.
If abnormal values continue to be measured despite performing zero calibration, a CO2 calibration will be conducted.
Zero calibration aligns the "0" position like a ruler, while CO2 calibration aligns the "38" position like a ruler (5% of gas corresponds to 5% of atmospheric pressure, which is 38 for a pressure of 760 mmHg).
For any further inquiries or questions regarding these procedures, please contact Capnotrainer Japan.